- Stable within ±0.01% per year
- 18 Models values from 1 µH to 10 H
- Low, known temperature coefficient
- 3 year warranty included
- Calibrazione accreditata ISO-17025 inclusa
- Errori di connessione ridotti al minimo
- Toroidale - privo di campi esterni
The GenRad 1482 Standard Inductors are the standard of choice in metrology labs. Used today by national metrology institutes and primary standards bodies around the world, these inductors have no peer or equivlent.
IET continues to produce the 1482 standard inductors to the same exacting specs as General Radio. These standard inductors are accurate, highly stable standards of self inductance for use as a low frequency reference of working standard in the laboratory.
Records extending over 30 years, including those of standard inductors that traveled to national laboratories in several countries for calibration, show long-term stability well within ±0.01%.
Each standard inductor is a uniformly wound toroid on a ceramic core. It has a negligible external magnetic field and hence essentially no pickup from external fields. The inductor is resiliently supported in a mixture of ground cork and silica gel, after which the whole assembly is cast with potting compound into a cubical aluminum case.
Sizes of 500 µH and above have three terminals, two for the inductor leads and the third connected to the case, to provide either a two- or three terminal standard. The 50-, 100-, and 200 µH sizes have three additional terminals for the switching used to minimize connection errors.
For comparing other inductors with these standards, the 1689 or 1693 Digibridge is recommended.
IET continues to produce the 1482 standard inductors to the same exacting specs as General Radio. These standard inductors are accurate, highly stable standards of self inductance for use as a low frequency reference of working standard in the laboratory.
Records extending over 30 years, including those of standard inductors that traveled to national laboratories in several countries for calibration, show long-term stability well within ±0.01%.
Each standard inductor is a uniformly wound toroid on a ceramic core. It has a negligible external magnetic field and hence essentially no pickup from external fields. The inductor is resiliently supported in a mixture of ground cork and silica gel, after which the whole assembly is cast with potting compound into a cubical aluminum case.
Sizes of 500 µH and above have three terminals, two for the inductor leads and the third connected to the case, to provide either a two- or three terminal standard. The 50-, 100-, and 200 µH sizes have three additional terminals for the switching used to minimize connection errors.
For comparing other inductors with these standards, the 1689 or 1693 Digibridge is recommended.
SPECIFICAZIONE DI BASE (le specifiche complete possono essere trovate sulla scheda tecnica)
Calibrazione: un certificato di calibrazione accreditato da A2LA a ISO-17025 viene fornito con ciascuna unità, fornendo valori misurati di induttanza a 100, 200, 400 e 1000 Hz, con le condizioni di prova e il metodo di misurazione specificati. Questi valori sono ottenuti mediante confronto, con una precisione, tipicamente, migliore dello 0,005%, con standard i cui valori assoluti, riconducibili al Sistema Internazionale di Unità (SI), sono noti con una precisione tipicamente di (0,02% + 0,1μH) a 100 Hz; (0,1% + 0,1 μH per il 1482-B).
Stabilità : il cambiamento di induttanza è inferiore allo 0,01% all'anno.